NAME mac_hdw - macro-hardware facility
DESCRIPTION
The macro-hardware facility links user-defined macro
functions with spec's built-in C code for motor,
counter and MCA control, providing a simple method to control
hardware for which spec doesn't have built-in
support. (MCA support added in release 5.09.03-1.)
The facility also provides a simplified method for
implementing calculational pseudomotors. These pseudomotors
correspond to motions formed by a combination of real motors.
For example, the motions of a slit's gap and offset are derived
from the real motions of the slit's blades, or a table's height,
pitch and roll are derived from the real motions of the table
legs.
The macro counter implementation also includes an optional
polling feature, to accommodate hardware devices that have a run
cycle that may take longer than the timing cycle controlled by
spec's configured master timer (as of
spec release 5.08.03-4).
For regular macro-hardware motors, counters and MCA devices,
there are three possible macro functions that can be user
defined. During start up and on the reconfig
command, for each macro-hardware "controller" unit and
for each macro-hardware motor and counter channel assigned in the
hardware config file, spec will call the
"configuration" user-defined macro function. For each
low-level command needed to implement a motor, counter or MCA
control function, spec will make calls to the
"command" user-defined macro function. Finally,
spec will call the "parameter" user-
defined macro function to set or get values for associated motor,
counter and MCA parameters.
Calculational pseudomotors are pseudomotors whose positions
depend on the positions of real motors. For such a macro-
hardware pseudomotor, spec will call a
"calculation" user-defined macro function that will
calculate the position of each pseudomotor based on the real
motor positions or will calculate the real motor positions based
on a target pseudomotor position. There is no
"command" macro function for calculational
pseudomotors.
To enable a macro motor or counter, configure a macro
controller on the Devices screen of configuration editor,
as follows:
| -
MOTORS DEVICE ADDR <>MODE NUM <>TYPE
YES motxx - 5 Macro Motor
SCALERS DEVICE ADDR <>MODE NUM <>TYPE
YES cntxx - 5 Macro Counter
YES cntzz - 5 Macro Counter/Timer
|
Here, the motxx, cntxx or
cntzz entries for DEVICE are arbitrary names, but
will be used by spec as a prefix for the macros
associated with the particular macro-hardware controller.
To enable a macro-hardware MCA, configure the controller on
the MCA-like configure screen as follows:
| -
MCA-like DEVICE ADDR <>MODE <>TYPE
0 YES mac_mca - Macro MCA
|
|
The ADDR field is optional in all cases, but may
contain a string value. For motors and counters, the value can
be retrieved within spec using
motor_par(mne,
"address")
or
counter_par(mne,
"address")
where mne is the mnemonic of any motor
or counter channel associated with the controller (as of
spec release 5.06.04-4). For MCA devices, use
mca_par("address")
or
mca_spar(num,
"address")
where num is the MCA number. In
addition, within the macro functions described below, the value
will assigned to a local variable named
prefix_ADDR (as of spec
release 5.06.04-8).
Nonstandard optional controller parameters, entered and
modified using the p command on the Devices
screen and stored in the config file prefixed by the
string CONPAR:, are accessible within the macro
hardware functions as elements of the associative array
prefix_CONPAR[], where the array elements are
indexed by the parameter name. In addition, the parameters for
motors and counters are also available using
motor_par(mne,
par)
or
counter_par(mne,
par)
where mne is the mnemonic of any motor
or counter channel associated with the controller and
par is the parameter name (as of
spec release 5.08.05-1). For MCA devices, use
mca_par(par)
or
mca_spar(num,
par)
where num is the MCA number.
Individual motor and counters are assigned to macro-hardware
controllers on the Motor and Scaler screens,
respectively. On the Motor screen, select
MAC_MOT as the controller type. For ordinary
counters, choose MAC_CNT as the controller type on
the Scaler screen. For polled counters (introduced in
spec release 5.08.03-3), choose the
MAC_CNTP controller type.
The Macro Functions
For standard macro motors and macro counters, the three
possible user-defined macro functions have names formed by
prepending the prefix defined in the config file to
_config(), _cmd() and
_par(). For calculational pseudomotors, a
_calc() function is needed rather than the
_cmd() function.
For all the macro functions, the first argument is the motor
or counter number if the call applies to a single channel or the
string ".." if the call applies to all channels
associated with the motor or counter controller. For MCA
devices, the first argument is the MCA number. The second
argument is a string that contains a key specific to the command.
Remaining arguments, if any, contain parameters.
As explained in the previous section, if an optional
ADDR string is included in the config file
for the associated controller, the string will be assigned to a
variable named prefix_ADDR. Likewise,
nonstandard optional parameters configured for the associated
controller are assigned to an associative array variable named
prefix_CONPAR. The variables are only
visible within the macro functions.
To send an error back to spec from the macro
functions, return the special string ".error.".
Set the spec DEBUG level to 128 to display
the macro function calls, which may help make clear when and in
what order the various macro functions are called.
The _config() Function
The _config() function is called after reading
the config file. On startup, the function is called after
the initial command files have been read, so that macros defined
in the initial command files can be used to set up the macro
hardware. For regular macro motors and counters, this function
is optional and need not exist. The function is required for
calculation pseudomotors to configure the dependencies between
real and pseudomotors. The function is also required for MCA
devices to configure the maximum number of channels and the
native data type.
For motor and counter controllers, the _config
macro function is called as follows:
prefix_config("..", "ctrl",
p1, p2) - Called
for each macro-hardware controller unit. The parameter
p1 is the unit number of the controller,
while p2 is the number of channels set as NUM
on the Devices screen of the configuration editor. If the
function returns the string ".error.",
spec will consider the controller unresponsive
and won't call the macro functions for the associated channels.
prefix_config(mne, "mot",
unit, module,
chan) - Called for each macro motor
channel, where
mne is the motor mnemonic,
unit is the unit number of the associated
controller, module is the optional module
number and chan is the channel number. For
calculational pseudomotors, this call must return a string
containing a list of mnemonics for the real motors on which the
pseudomotor depends. If the function returns the string
".error.", spec will consider the
channel unusable.
prefix_config(mne, "cnt",
unit, 0, chan)
- Called for each macro counter channel, where
mne is the counter mnemonic,
unit is the unit number of the associated
controller, and chan is the channel number.
The fourth argument is currently always zero. If the function
returns the string ".error.", spec
will consider the channel unusable.
prefix_config(num,
"mca") - Called for each macro MCA, where
num is the MCA number. This function must
return a string with two arguments in any order: the maximum
number of channels for the MCA and the native data type. The
data type is one of the words byte,
ubyte, short, ushort,
long, ulong, float or
double. An example return string is
"4096 ulong". If the function returns the
string ".error.", spec will
consider the MCA unit unusable.
For motors and counters, the unit and channel numbers are
assigned in the configuration editor just as with other motors
and counters. The optional module number became available in
spec release 5.04.03-2. Unit numbers are
assigned consecutively to each controller type. That is, the
first macro motor controller listed is unit zero, as is the first
macro counter controller, both independent of other controller
types. Counter unit and channel numbers are assigned explicitly
on the Scaler screen of the configuration editor. Motor
unit, module and channel numbers can be explicitly assigned in
the unit/channel field of the configuration editor. If left
blank, the channels numbers are assigned automatically in
consecutive order.
For macro hardware designed for general purpose applications,
these calls to the _config() macro function can be
used to set up the rest of the macro interface. For simple
applications, this function won't be required for motors and
counters.
MOTORS
The _cmd() Function For Motors
The _cmd() function is called to control regular
macro motors. There are many more command keys in the list below
than any particular macro motor would need. Only the command
keys that are relevant to the particular application should be
included in the user-defined macro function.
The syntax of the function call is:
prefix_cmd(mne,
key [, p1 [,
p2]] [,unit])
Called by the C code for all operations related to motor
control. mne is the string ".."
for keys that apply to all motors. mne is
the motor number for keys that apply to individual motors. Each
key is only called one way or the other. key
is a string containing the particular command.
p1 and p2, if present,
are parameters related to the command. If the second argument is
the string ".." the unit
parameter will be included and specifies the unit number for
which the command applies (as of spec release
5.07.03-3).
In the following, the phrase "sent when changed"
means the _cmd() macro function is only called with
the given key before the first applicable move command or home
command after spec reads the config file,
either on startup or after a reconfig command
(included in the config macro) or after the
associated parameter has been changed using the motor_par()
function.
"base_rate" Sent when changed. The
p1 parameter contains the base rate in units
of Hz.
"slew_rate" Sent when changed. The
p1 parameter contains the slew rate in units
of Hz.
"acceleration" Sent when changed. Also
called if the base rate or slew rate have changed, since some
motor controllers need to be told to recalculate acceleration
ramps if the velocity parameters change. The
p1 parameter contains the acceleration time
in units of milliseconds. The p2 parameter
contains the acceleration in units of steps per second per
second.
"home_base_rate" - Sent when changed. The
p1 parameter contains the home base rate in
units of Hz.
"home_slew_rate" - Sent when changed. The
p1 parameter contains the home slew rate in
units of Hz.
"home_acceleration" - Sent when changed.
The
p1 parameter contains the home
acceleration time in units of milliseconds. The
p2 parameter contains the home acceleration
in units of steps per second per second.
"preread_all" - Sent prior to a possible
read of all the motors. Note, depending on the configured
hardware read modes for the motors, there may be no subsequent
commands to read a motor associated with this controller. Either
"preread_all" or "preread_one" (below),
but not both, will be called prior to the "position"
call below.
"preread_one" - Sent prior to reading an
individual motor. Either
"preread_all" (above) or
"preread_one", but not both, will be called prior to
the "position" call below.
"position" - For this key, the macro
function must return the current motor position in dial units.
This call will be preceded by a call with a key of either
"preread_one" or "preread_all".
The macro-hardware motors use natural units for positions.
For such motor controllers, spec only uses the
steps-per-degree parameter in the config file to determine
the precision of the position values. The precision is
determined by the magnitude of the parameter. For example, a
value of 1000 will cause spec to round to the
nearest 0.001, while a value of 5000 will cause
spec to round to the nearest 0.0002. The
rounding is performed on the value returned by the
"position" key.
"set_position" - Sent to set the current
dial position of the macro motor. The parameter
p1 contains the position in dial units.
"prestart_all" - For regular moves, sent if
any motors associated with the macro motor controller need to be
moved.
"prestart_one" - For regular moves and
homing moves, sent for each motor that needs to be moved. For
regular moves, a call of
"prestart_all" comes first.
"magnitude" - For regular moves, sent with
the magnitude of the move in dial units. The parameter
p1 contains the position in dial units and
includes the sign of the move. The magnitude is also included
with the "start_one" key (below).
"start_one" - Sent to start a regular move
for one motor. The parameter
p1 contains the
target position in dial units to accommodate a controller that
requires absolute positions. The parameter
p2 contains the magnitude of the move in dial
units to accommodate a controller that requires relative
positions.
"start_all" - Sent after all
"start_one" commands to accommodate controllers that
use a simultaneous start.
"get_status" - Sent to get the move and
limit status of individual motors. If the motor is moving, the
macro function must return a value with bit 0x02 set. If the low
limit is active, the return value must have bit 0x04 set. If the
high limit is active, the return value must have bit 0x08 set.
Setting bit 0x10 indicates an "emergency stop" and
setting bit 0x20 indicates a motor fault (as of
spec release 5.07.04-4), both which currently
result in similar behavior to when a limit is hit. Otherwise,
the macro function must return a value of zero.
"flush_all" - Sent before the
"hard" position synchronization that occurs on startup
and before and after reading the config and
settings files on a
reconfig command.
"flush_one" - Sent for each motor after the
"flush_all" key above and before a
"get_status" during the position synchronization.
"abort_one" - Sent for each active motor
when motors are halted, normally either by a
^C from
the keyboard or by a stop() command.
"abort_all" - Sent to each macro motor
controller that has busy motors when motor are halted. The key
is sent after the
"abort_one" keys are sent. This
command accommodates controllers that allow a single command to
halt all its associated motors.
"search" - Sent to initiate a home or limit
search. A
"prestart_one" call will precede this
call. The parameter p1 indicates the type of
search as follows.
"home" - Find the home position as
appropriate.
"home+" - Find the home position by moving
in the positive direction.
"home-" - Find the home position by moving
in the negative direction.
"lim+" - Find the positive limit.
"lim-" - Find the negative limit.
The parameter p2 contains the position
in dial units that corresponds to the home or limit switch.
"diff_position" - Sent if the motor is
configured for a settle time. To configure a motor for settle
time, both the DC dead-band and the DC settle-time parameters
have to be set (usually from the first optional motor parameter
screen of the configuration editor - get there by typing an
m from the primary motor screen). The settle-time
parameter is in seconds. spec will wait for at
least as long as the settle time before treating a move as
complete. In addition, spec will wait until the
value returned by this call is less than the dead band, but for
no longer than five seconds. The preferred units for dead band
are steps, but it is only necessary that the units in the
config file agree with the units returned by this call.
Backlash is handled as two separate move commands. If the
macro function will take care of backlash, set the backlash
parameter to zero in the config file.
The default behavior with respect to reading the motor
position is to only ask for the motor position from the
controller during position synchronization or at the end of a
move. The motor parameter "hardware read mode" (on the
second optional motor parameter screen of the configuration
editor) can be set to require spec to ask for
the position before each move and/or for every
read_motors() call from user level. The hardware
read mode can also be set so that position discrepancies are
always silently resolved in favor of the value returned by the
controller (or macro function).
A minimal implementation would likely recognize the keys
"start_one", "get_status",
"position" and "set_position". An
example that does nothing useful follows:
| -
def motxx_cmd(mne, key, p1, p2) '{
global demo_pos[]
if (key == "set_position") {
demo_pos[mne] = p1
return
}
if (key == "position") {
return(demo_pos[mne])
}
if (key == "start_one") {
demo_pos[mne] = p1
return
}
if (key == "get_status") {
return(0)
}
}'
|
The _par() Function For Motors
The _par() macro function is called when various
motor parameters are set, and when the motor_par()
function is used to retrieve a user-defined parameter. The
mne argument will always be a motor number
and never the string ".." that is used with the
other user-defined macro-hardware functions.
The function will be called as:
prefix_par(mne,
key, "get") - The function should
return a value for the parameter named as
key
for motor number mne. The macro function
will never be called to get a parameter value when
key is a parameter name that is built into
the spec C code.
prefix_par(mne,
key, "set", p1 [,
p2]) - Called when various
parameters change their value, as described below.
The built-in parameter names are as follows. The
_par() function will never be called with
"get" for these parameters, as their values are
maintained internally.
The first set below are called only when
motor_par() is executed from user-level.
"acceleration" p1
contains the acceleration time in milliseconds.
p2 contains the acceleration value in steps
per second per second.
"backlash" p1
contains the new backlash value in steps.
"backlash_rate" p1
contains the new backlash rate in Hz.
"base_rate" p1
contains the new base rate in Hz.
"disable" p1 contains
1 or 0, depending on whether the motor was disabled or un-
disabled (available as of spec release
5.06.03-8).
"slew_rate" or "velocity"
p1 contains the new slew rate
in Hz.
"step_size" p1
contains the new step-size parameter.
The following two keys are called when an associated function
is executed from user level.
"limits" p1 contains
the low limit in dial units. p2 contains the
high limit in dial units. Called when the user-level
set_lim() function is executed.
"offset" - Called when the user-level
chg_offset() function is executed.
p1 contains the offset in user units.
The following optional motor parameters generate a call to
the user-defined macro-hardware function when the values are read
from the config file and when the values are set with the
motor_par() function. See the
motors help file for additional
information on the parameters.
"home_slew_rate"
"home_base_rate"
"home_acceleration"
"dc_dead_band"
"dc_settle_time"
"dc_proportional_gain"
"dc_derivative_gain"
"dc_integral_gain"
"dc_integration_limit"
"dc_following_error"
"dc_sampling_interval"
"encoder_step_size"
"step_mode"
"slop"
"read_mode"
"deceleration"
"torque"
"misc_1"
"misc_2"
"misc_3"
"misc_4"
"misc_5"
"misc_6"
There are a number of valid arguments to
motor_par() which will not generate a call to the
_par() macro function at all. These include
"disable", "unit",
"channel", "responsive",
"controller", "device_id",
"active", "status",
"config_step_size",
"config_acceleration",
"config_velocity", "config_base_rate",
"config_backlash", "low_limit" and
"high_limit".
Arguments to motor_par() that are not recognized
by the built-in C code will be passed on, as is, to the
_par() user-defined macro function.
Calculational Pseudo Motors
For calculational pseudomotors, two macro functions must be
be provided with names formed by prepending the prefix from the
config file to _config() and
_calc().
The _config() function, when called with the key
equal to "mot", must return a string that contains a
space-delimited list of mnemonics for the real motors on which
motor mne depends.
The _calc() macro function will be called as
follows:
prefix_calc(mne,
mode) - When called with
mode equal to zero, the function should
assign a value to A[mne]
corresponding to the current position of the pseudomotor
mne. When called with
mode equal to one, the function should assign
a value to A[mne] corresponding
to the target position of the real motor mne.
When called to calculate the real motor positions for a move
(with mode equal 1), the function will first
be called with mne set to the string
"..", then called with each of the real motor
mnemonics as arguments, in turn. One can use the initial call to
calculate quantities that depend on the current positions of the
real motors before new values are assigned in subsequent calls.
The _calc() function should only include
commands to calculate pseudomotor positions from real motor
positions or vice versa. The function should not contain calls
to do hardware access. In fact, calls to the built-in functions
wait() or read_motors() will return
immediately if called from the _calc() macro
function, to avoid possible recursion as those built-in functions
can subsequently call the invoking _calc() macro
function.
Note, if the pseudomotor is disabled via
motor_par(mne, "disable", 1)
the calls to _calc() with mode
equal to 1 will be skipped. That will prevent new positions from
being calculated for the associated real motors.
With spec release 5.06.04-4, a special
motor_par() option called "chan0" is
available for macro motors. The command
|
will return the motor number of the macro motor in the first
channel of the same controller unit and module number of the
macro motor with mnemonic mne. This feature
allows for simplified implementation of general-purpose
calculational pseudomotors. For example, if there are parameters
associated with a group of motors, one copy of the parameters can
be assigned to the "chan0" motor and then accessed
by the related motors.
The following examples implements calculational pseudomotors
for a slit. The slit has two blades whose real motors have
mnemonics sl2t and sl2b (slit 2 top and
bottom). The pseudomotors are the slit gap and the slit offset
position with mnemonics sl2g and sl2o,
respectively.
The Devices screen of the configuration editor should
look as follows for the controller:
| -
MOTORS DEVICE ADDR <>MODE NUM <>TYPE
YES slit2 2 Macro Motor
|
|
The macros would be as follows:
| -
def slit2_config(mne, type, unit, module, chan) '{
if (type == "mot")
return "sl2t sl2b"
}'
def slit2_calc(mne, mode) '{
if (mode == 0) {
if (mne == sl2g)
A[mne] = A[sl2t] + A[sl2b]
if (mne == sl2o)
A[mne] = (A[sl2t] - A[sl2b])/2
} else {
if (mne == sl2b)
A[mne] = -A[sl2o] + A[sl2g]/2
if (mne == sl2t)
A[mne] = A[sl2o] + A[sl2g]/2
}
}'
|
|
The following make_slit_macs macro can be used
to generate macros such as the above.
| -
def make_slit_macs '{
local name, file
local l, r, g, o
file = getval("File for macros", "tty")
name = getval("Name of slit", "Slit1")
r = getval("Mnemonic for right/top slit", "s1r")
l = getval("Mnemonic for left/bottom slit", "s1l")
g = getval("Mnemonic for gap", "s1vg")
o = getval("Mnemonic for offset", "s1vo")
fprintf(file, "\n\
def %s_config(mne, type, unit, module, chan) \'{\n\
if (type == \"mot\")\n\
return \"%s %s\"\n\
}\'\n\
def %s_calc(mne, mode) \'{\n\
if (mode == 0) {\n\
if (mne == %s)\n\
A[mne] = A[%s] + A[%s]\n\
if (mne == %s)\n\
A[mne] = (A[%s] - A[%s])/2\n\
} else {\n\
if (mne == %s)\n\
A[mne] = -A[%s] + A[%s]/2\n\
if (mne == %s)\n\
A[mne] = A[%s] + A[%s]/2\n\
}
}\'\n", name,l,r, name, g,r,l, o,r,l, l,o,g, r,o,g)
}'
|
|
The following example implements a table-height pseudomotor
with mnemonic t1z that is the average height of the
three real motors t1f, t1b1 and
t1b2 that correspond to the table legs. When the
height is moved, each leg is moved by an amount equal to the
difference of the current height and the target height. The
current average height needs to be calculated from the current
real-motor positions before the new positions are assigned. The
feature where the _calc() function is called with
mne set to the string ".." before being
called with the real motor mnemonics is used to save the average
position in a global variable to be used in the subsequent calls.
| -
def tab1_config(mne, type, unit, module, chan) '{
global tab1_ave
if (type == "mot")
return "t1f t1b1 t1b2"
}'
def TE1_vert_jack_calc(mne, mode) '{
if (mode == 0) {
if (mne == t1z)
A[mne] = (A[t1f] + A[t1b1] + A[t1b2])/3
} else {
if (mne == "..")
tab1_ave = (A[t1f]+A[t1b1]+A[t1b2])/3
else if (mne == t1f)
A[mne] += A[t1z] - tab1_ave
else if (mne == t1b1)
A[mne] += A[t1z] - tab1_ave
else if (mne == t1b2)
A[mne] += A[t1z] - tab1_ave
}
}'
|
|
This last example shows how an energy pseudomotor can
be readily created that ties in with the standard
spec energy macros from the energy.mac
distribution file. Such a pseudomotor can then be scanned using
the standard motor scans. Note, the existing energy macros, such
as Escan, moveE, etc., will still work.
The _config() function below makes use of the
monochromator mnemonic conventions set in energy.mac.
| -
def Energy_config(mne, type, unit, module, chan) '{
if (type == "mot") {
if (mono_type == 1)
return(motor_mne(Mono))
if (mono_type == 2)
return("mono mon_y mon_z")
if (mono_type == 3)
return("monu mond montrav")
if (mono_type == 4)
return("monu mond montrav monoff")
}
}'
def Energy_calc(mne, mode) '{
if (mode == 0) {
calcE
A[energy] = hc_over_e / LAMBDA
} else {
calcM A[energy]
calcE
}
}'
|
|
The prefix Energy and motor mnemonic
energy are, as always, arbitrary, but must match the
configured device name on the Devices screen and motor
mnemonic on the Motor screen.
COUNTER/TIMERS
The _cmd() Function For Counters
The _cmd() macro function for counters is called
in response to spec's counter-related built-in
functions and commands: tcount(),
mcount(), wait(), stop(),
move_cnt, getcounts and
sync.
There are three possible counter channel configurations: a
master timer, a regular counter channel or a polled counter
channel. The function calls differ depending on the
configuration.
A master timer belongs to controller type "Macro-
Hardware Counter/Timer". The counter type is
MAC_CNT and the individual channel is identified as
either timer or monitor. For such a
channel, the "start_one" call will include a nonzero
parameter indicating the count mode as described below. Also,
"get_status" calls will be made to determine when
the count time has completed.
A regular counter channel belongs to controller type
"Macro-Hardware Counter" with counter type
MAC_CNT and function type counter.
The polled counter type also has controller type "Macro-
Hardware Counter" and function type counter,
but the counter type is MAC_CNTP. Polled counters
will receive "get_status" calls to see when the
counter is no longer busy.
The syntax of the function call is:
prefix_cmd(mne,
key [, p1 [,
p2]] [,unit])
Called by the C code for all operations related to
counter/timer control. mne is the string
".." for keys that apply to all counters.
mne is the counter number for keys that apply
to individual counters. Each key is only called one way or the
other. key is a string containing the
particular command. p1 and
p2, if present, are parameters related to the
command. If the mne argument is the string
".." the unit parameter will be
included and specifies the unit number for which the command
applies (as of spec release 5.07.03-3).
When the "start_all" and
"start_one" keys are sent as described below, the
count-mode parameter will be set as follows:
| -
0 - this channel is not the master timer
1 - count to a monitor preset - mcount()
2 - count to a time preset - tcount()
3 - just count until the counters are halted - move_cnt
|
|
Possible keys are as follows:
"prestart_all" Sent to the controller
(mne set to "..") before any
counters are started. p1 will contain the
count preset, either in seconds or monitor counts.
p2 contains the count mode (1, 2 or 3), as in
the above list. The fifth argument will contain the unit number
of the controller.
"start_one" Sent to start each counter at
the start of the counting period. The function will be called
for the regular counting channels before being called for the
master timer channel. p1 is the count time
in seconds if counting to time or in counts if counting to
monitor. If mne is the master timer,
p2 will contain 1, 2 or 3 to indicate the
counting mode. If mne is not the master,
p2 will be zero.
"get_status" Sent after the counters have
been started, but only to the master timer/counter, if there is
one, or to any counters configured as polled (with type
MAC_CNTP). Must return nonzero if the channel is
busy or zero if the counting is finished.
"counts" Called when
spec's built-in getcounts function
is executed. The macro function must return the current counts
for scaler channel number mne. Note, all the
scalers channels associated with real counters will have been
read before the macro function call, so the values is the
S[] counter array will be current for the non-macro
counters.
"halt_all" Sent to each macro counter
controller (mne set to "..")
when counting is halted. The key is sent before the
"halt_one" keys are sent. The third argument will
contain the unit number of the controller.
"halt_one" Sent to each active counter
when counting is halted. The p1 argument
will be zero if called at the end of normal counting and one if
called when ^C is typed at the keyboard or when a
stop() or sync command is entered or
after motors have stopped with a move_cnt.
Commands to halt counters are sent when a preset count time
elapses, if a ^C is typed from the keyboard, when a
stop() or sync command is encountered
from user-level or after motors have stopped when the
move_cnt command is used.
The _par() Function For Counters
There are currently no built-in keywords that produce calls
to the _par() macro function for counters. The
macro function will only be called when spec's
user-level counter_par() function is called to set
or retrieve an otherwise unrecognized parameter.
The function will be called as:
prefix_par(mne,
key, "get") - The function should
return a value for the user-defined parameter named as
key for counter number
mne.
prefix_par(mne,
key, "set", p1)
- Called to set the user-defined parameter
key to p1 for counter
number mne.
The mne argument will always be a counter
number and never the string "..".
Built-in arguments to counter_par() that can't
be passed to a user-defined macro function include
"disable", "unit",
"channel", "responsive",
"controller", "device_id",
"scale", "monitor", and
"timer".
MCAs
The _cmd() Function For MCA Devices
The _cmd() macro function for MCA devices is
called in response to spec's MCA and counting-
related built-in functions and commands: mca_get(),
mca_put(), tcount(),
mcount(), move_cnt,
wait(), stop() and sync.
The _cmd() macro function will also be called in
response to the standard mca_par() commands
"clear", "run", and
"halt".
The syntax of the function call is:
prefix_cmd(num,
key [, p1 [,
p2 [, data]]])
Called by the C code for operations related to MCA control.
num is the MCA number.
key is a string containing the particular
command. p1 and p2, if
present, are parameters related to the command. For the
mca_get() and mca_put() functions,
data will be a data array for sending or
receiving the MCA data.
None of the keys are mandatory. Possible keys are as
follows:
"clear" - Sent in response to
mca_par("clear") and before "run" when
"auto_clear" mode is enabled. There are no
additional parameters.
"run" - Sent in response to
mca_par("run") and at the start of counting
associated with the standard tcount(),
mcount() and move_cnt commands when
"auto_run" mode is enabled. If
"soft_preset" mode is enabled,
p1 will be the counting preset (seconds or
monitor counts), otherwise p1 will be the
value set with the mca_par() "preset"
option. p2 will be the count mode, as
follows:
|
-
-
1 - called by
tcount()
2 - called by mcount()
3 - called via move_cnt
4 - called by mca_par("run")
|
-
"halt" - Sent in response to
mca_par("halt") and at the end of counting
associated with the standard tcount(),
mcount() and move_cnt commands when
"auto_run" mode is enabled. Not called at the
normal end of counting if "soft_preset" is enabled
or if the mca_par() "preset" value is
nonzero. The p1 argument will be zero if
called at the end of normal counting and one if called when
^C is typed at the keyboard or when a
stop() or sync command is entered or
after motors have stopped with a move_cnt.
"get_status" - Called during counting. A
return value of one will indicate the MCA is still busy.
"read" - Sent to read data.
p1 and p2 are the first
and last channels to be read, respectively, reflecting the
optional first and last channel arguments to
mca_get(). data is a data array
of the native type (configured with the _config()
call) where the data should be written. The optional return
value is the number points actually read. The value will be used
by mca_get() to determine how many points to copy to
the returned array. If no value is returned,
mca_get() will use the requested number of points.
"write" - Sent to write data.
p1 and p2 are the first
and last channels to be write, respectively, reflecting the
optional first and last channel arguments to
mca_write(). data is a data
array of the native type (configured with the
_config() call) from where the data can be read.
The optional return value is the number points actually written.
The _par() Function For MCA Devices
The _par() macro function is called when various
MCA parameters are set, and when the mca_par() or
mca_spar() functions are used to retrieve a user-
defined parameter. The num argument is the
MCA number.
The standard mca_par() commands
"info", "chans",
"max_chans", "auto_run",
"soft_preset", "auto_clear",
"native_type", "preset", and
"chan#", are handled by
spec's built-in code and will not generate calls
to the _par() macro function. The standard
mca_par() "disable" call will be passed
to the _par() function when setting or clearing the
disabled state.
The function will be called as:
prefix_par(num,
key, "get") - The function should
return a value for the user-defined parameter named as
key for MCA number num.
prefix_par(num,
key, "set", p1)
- Called to set the user-defined parameter
key to p1 for MCA number
num.
MCA Counting Mode Summary
The mca_par() "auto_run" mode must
be enabled for the standard spec counting
commands to control the MCA. If the mca_par()
"soft_preset" mode is enabled, when the
_cmd() function is called with the
"run" argument, p1 will be the
count time, seconds when counting to time and counts when
counting to monitor. When in this mode, the macro hardware MCA
will be polled to determine when it is finished. If the MCA
takes longer to count to the preset than the master timer,
spec will wait until the MCA has finished.
Unless the counting is aborted, The "halt" command
will not be called, as the macro hardware MCA is expected to
count until it reaches its programmed preset value. The
p1 argument will be zero for powder-mode
counting with the move_count command, and the MCA
will be sent a "halt" command when
spec senses counting has finished due to the
designated motor having completed its trajectory.
If "soft_preset" is disabled, when the
_cmd() function is called with the
"run" argument p1 will be the
value set using mca_par() "preset", or
zero if the value has not been set. If the value is nonzero,
spec will poll the MCA until it returns a non-
busy status. Thus one can have the macro hardware MCA count to a
different preset than the master timer. If the preset value is
zero, the MCA will be sent a halt command when the master timer
reaches the end of the count interval.
In all cases, a count abort will generate a
"halt" command.