NAME MPII - Canberra Multiport II USB/Socket MCA
INTRODUCTION
The Canberra Multiport II (MPII) is a NIM module containing
from one to six independent multichannel analyzer (MCA) devices.
The MPII can be used with spec over both socket
and USB interfaces. Each MCA device can be configured for up to
16,384 channels and can be operated in either pulse-height
analysis (PHA) or multichannel scaling (MCS) mode.
CONFIGURATION
Each MPII module interfaced to spec over a socket interface
can be uniquely identified by its assigned IP address. USB units
cannot be assigned addresses, so if more than one MPII is used
with spec over USB, assignment of specific modules is done by
including the module's unique serial number in
spec's hardware config file. If the
serial number is entered as zero, spec will
simply assign the MCA units in the order in which they are found.
In the edconf configuration editor, possible MPII
configurations include:
| -
MCA- and CCD-Acquisition Type Device Configuration
MCA-like DEVICE ADDR <>MODE <>TYPE
0 YES host1:6000 Canberra Multiport II (Socket)
1 YES 192.168.1.5 Canberra Multiport II (Socket)
2 YES 12345 Canberra Multiport II (USB)
3 YES 0 Canberra Multiport II (USB)
4 YES 0 Canberra Multiport II (USB)
|
The first entry configures a socket-based MCA using a resolvable
host name and an explicit port number. Note, the default port
for the MPII is 6000, and doesn't need to be specified. The
second entry specifies an IP address without the optional port
number. Three USB configurations follow, the first with an
explicit module serial number. The last two entries are also for
USB units, but with the serial number set to zero. For those two
units, it won't be possible to know in advance which MPII will be
assigned to MCA unit number 3 and which to MCA unit number 4.
ADDRESSING
The spec mca_get(),
mca_put() and mca_par() functions work
with the default MCA unit. If there is only one MCA unit
configured, that unit is the default. Otherwise, the
mca_sel() function can be used to select the default
unit number. The mca_spar(),
mca_sget() and mca_sput() take an
initial argument that specifies a specific MCA unit number,
independent of the default MCA unit number. See the
mca help file for more details.
Because each MPII can have up to 6 independent MCA devices,
the MCA selection syntax for the commands
mca_spar(), mca_sget(), and
mca_sput() is expanded. To select a specific MCA
device on an MPII, address arguments may be given as
0.1, 0.2, ..., 0.6, where
the first digit is the MCA unit number from the config
file, and the digit right of the decimal point is the MCA device,
where the devices are numbered from 1 to 6. The syntax
"0:1", "0:2", ..., "0:6"
is also recognized, where the arguments are strings.
For writing data to the MCA with mca_sput() and
for setting parameter values with mca_spar(), if no
subaddress is specified, the command will apply to all of the up-
to-six MCA devices available. That is, the same data or
parameters can be written to all MCA devices on a Canberra module
with one command. For reading data or reading parameters values,
only one MCA device can be accessed since only one set of data or
a single value can be returned. Thus, a subaddress needs to be
specified for commands that return data if there is more than one
MCA device on the MPII unit.
If no subaddress is specified for any of the commands,
subaddress 1 is assumed. If the subaddress doesn't specify a
valid MCA device for any command, subaddress 1 is still assumed.
If there is only one MCA device on the unit, no subaddress is
necessary. If the MPII is the default MCA and there is only one
MCA device, mca_get(), mca_put() and
mca_par() functions can be used unambiguously.
Although the mca_sel() function can select the
default unit, it is not possible to assign a default subaddress.
FUNCTIONS
The standard MCA commands supported by spec
are described in the
mca help file. Refer to that
document for more details on usage of the mca_get(),
mca_sget(), mca_put() and
mca_sput() functions. Also, refer to the
mca help file for a description of
the standard mca_par() and mca_spar()
options "info", "chans",
"max_chans", "disable",
"auto_run", "soft_preset",
"auto_clear" and "native_type".
The default settings for the MPII are for
"auto_run" and "soft_preset" to be
disabled and "auto_clear" to be enabled. These
default settings are restored on start up and hardware
reconfiguration. When "auto_run" and
"soft_preset" modes are in effect, the available
MPII devices will be programmed with a time preset when counting
to time and with the gross-counts preset when counting to
monitor. When counting to time, the mode will be real-time or
live-time depending on whether the MPII is in real- or live-
preset mode. That mode is set by using mca_par()
with "preset_live" or "preset_real".
Of the preset modes, real, live and gross counts, only one
can be active at a time.
Note, of the parameters described below, the following are
saved in the state file and will maintain values between
spec invocations: "xstart_enable",
"xstop_enable", "mcs_xsync",
"mcs_start_chan", "mcs_stop_chan",
"overflow_enable", "preset_passes",
"preset_real", "preset_live" and
"preset_counts". When starting fresh, values will
be restored to defaults. All other MPII parameters are retained
in the module's flash memory, and spec will use
those values, including the values for "npts" (or
"adc_gain").
Additional options for the MPII are described below. The
string arguments are case insensitive.
Basic Operation
mca_par("clear") - Clears the current data
and the elapsed real-time, live-time, gross-counts and sweep-
counts registers.
mca_par("run") - Sets the current
acquisition parameters and starts acquisition. If
"auto_clear" mode is set, clears the existing data
first. Otherwise, if the device hasn't been explicitly cleared,
new data will be added to the existing.
mca_par("halt") - Halts acquisition.
mca_par("pha") - Sets pulse-height analysis
mode.
mca_par("npts" [, value])
- With no arguments, returns the current number of points.
Otherwise, sets the number of points by setting the ADC gain to
the
value specified. Synonymous with
"adc_gain". Allowed values are 256, 512, 1024,
2048, 4096, 8192 and 16384.
mca_par("preset_real" [,
value]) - With no arguments, returns
the current PHA-mode real (true) time preset in seconds.
Otherwise, sets the real-time preset to
value.
mca_par("preset_live" [,
value]) - With no arguments, returns
the current PHA-mode live-time preset in seconds. Otherwise,
sets the live-time preset to
value.
mca_par("preset_counts" [,
value]) - With no arguments, returns
the current PHA-mode gross-counts preset. Otherwise, sets the
gross-counts preset to
value.
mca_par("overflow_enable" [,
value]) - If no arguments, returns 0
if channel overflows are ignored and 1 if channel overflows halt
acquisition. Otherwise sets the overflow-enable mode according
to
value.
mca_par("xstart_enable" [,
value]) - With no arguments, returns
nonzero if the
XINPUT signal is to be used to start
acquisition. Otherwise sets external-start enable mode according
to value.
mca_par("xstop_enable" [,
value]) - With no arguments, returns
nonzero if the
XINPUT signal is to be used to stop
acquisition. Otherwise sets external-stop enable mode according
to value.
mca_par("save_interval" [,
value]) - With no arguments, returns
the interval in minutes at which the Mulitport II saves its
parameters to built-in flash memory. Otherwise sets the interval
as specified by
value. Allowed values are
from 1 to 60 minutes.
Status
mca_par("dump") - Displays all the
parameters and settings.
mca_par("elapsed_real") - Returns the
elapsed real (or true) time in seconds.
mca_par("elapsed_live") - Returns the
elapsed live time in seconds.
mca_par("elasped_counts") - Returns the
total elapsed counts.
mca_par("start_time") - Returns the
starting time of the most recent acquisition in seconds since
January 1, 1970, 00:00:00 GMT. The value can be converted to an
ASCII string using spec's
date()
function.
mca_par("stop_time") - Returns the stopping
time of the most recent acquisition in seconds since January 1,
1970, 00:00:00 GMT. The value can be converted to an ASCII
string using spec's
date()
function.
mca_par("stop_event") - Returns a code with
bits indicating the type of event that halted the most recent
acquisition as follows:
|
-
-
0x00 None
0x01 Real-time preset reached
0x02 Live-time preset reached
0x04 Total-counts preset reached
0x08 External input
0x10 MCS-preset passes reached
0x20 Power lost
0x40 Channel overflow
|
-
mca_par("overflow_chan") - If acquisition
was stopped by an overflow, returns the channel number that
overflowed.
mca_par("suspend_active") - Returns nonzero
if the XSUSPEND input is active, otherwise returns zero.
mca_par("collecting") - Returns nonzero if
the MCA device is busy collecting data.
mca_par("waiting_xinput") - Returns nonzero
if the MCA is waiting for the XINPUT to become active.
mca_par("start_delay") - Returns the
internal delay time in seconds for the MCA to begin the most
recent acquisition. This is a measure of latency within the MCA
device.
mca_par("changer_state") - Returns the
sample-changer state, where nonzero indicates the sample changer
is active.
External Connection Configuration
mca_par("xcoinc_polarity" [,
"low"|"high"|0|1]) - With no arguments, returns 0 (for
active low) or 1 (for active high) to indicate the polarity of
the
XCOINC input signal. Otherwise, sets the
polarity.
mca_par("xdt_polarity" [, "low"|"high"|0|1])
- With no arguments, returns 0 (for active low) or 1 (for
active high) to indicate the polarity of the
XDT
pileup-reject dead-time input signal. Otherwise, sets the
polarity.
mca_par("xlg_polarity" [, "low"|"high"|0|1])
- With no arguments, returns 0 (for active low) or 1 (for
active high) to indicate the polarity of the
XLG
pileup-reject linear-gate output signal, which is active when the
ADC acquires an input pulse. Otherwise, sets the polarity.
mca_par("xrej_polarity" [, "low"|"high"|0|1])
- With no arguments, returns 0 (for active low) or 1 (for
active high) to indicate the polarity of the
XREJ
pileup-reject input signal. Otherwise, sets the polarity.
mca_par("xcollectstatus_polarity" [,
"low"|"high"|0|1]) - With no arguments, returns 0 (for
active low) or 1 (for active high) to indicate the polarity of
the
XCOLLECTSTATUS external acquire status output
signal. Otherwise, sets the polarity.
mca_par("xinput_polarity" [,
"low"|"high"|0|1]) - With no arguments, returns 0 (for
active low) or 1 (for active high) to indicate the polarity of
the
XINPUT external start/stop and sample-changer
ready input signal. Otherwise, sets the polarity.
mca_par("xoutput_polarity" [,
"low"|"high"|0|1]) - With no arguments, returns 0 (for
active low) or 1 (for active high) to indicate the polarity of
the
XOUPUT advance-sample-changer output signal.
Otherwise, sets the polarity.
ADC Configuration
mca_par("adc_gain" [,
value]) - With no arguments, returns
the ADC gain. Otherwise, sets the ADC gain (and the number of
points) to the
value specified. Synonymous
with "npts". Allowed values are 256, 512, 1024,
2048, 4096, 8192 and 16384.
mca_par("adc_LLD" [,
value]) - With no arguments, returns
the current setting of the low-level discriminator as a percent.
Otherwise, set the low-level discriminator to the
value given as a percent from 0 to 100.
mca_par("adc_ULD" [,
value]) - With no arguments, returns
the current setting of the upper-level discriminator as a
percent. Otherwise, set the upper-level discriminator to the
value given as percent from 0 to 110.
mca_par("adc_zero" [,
value]) - With no arguments, returns
the current zero-discriminator setting as a percent. Otherwise,
sets the zero discriminator to the
value
given as a percent from -2.5 to 2.5.
mca_par("adc_coincidence" [,
value]) - With no arguments, returns
the current ADC coincidence mode, where 0 indicates normal and 1
indicates anti-coincidence mode. Otherwise, sets the ADC
coincidence mode as specified by
value.
mca_par("adc_high_pulse_action" [,
value]) - If no arguments, returns
zero if input pulses above the upper-level discriminator are
rejected and nonzero if such pulses are added to the last
channel. Otherwise sets the high-pulse action based on
value.
mca_par("adc_coinc_trigger" [,
value]) - If no arguments, returns
zero if the coincidence trigger is level sensitive and nonzero if
the trigger is edge sensitive. Otherwise sets the trigger mode
according to
value.
Multichannel Scaler
mca_par("mcs") - Set multi-channel scaling
mode.
mca_par("dwell" [,
value]) - With no arguments, returns
the MCS-mode dwell time in seconds. Otherwise, sets the dwell
time to
value. A negative argument selects
external dwell.
mca_par("mcs_start_chan" [,
value]) - With no arguments, returns
the start channel in MCS mode. Otherwise, sets the MCS start
channel to
value.
mca_par("mcs_stop_chan" [,
value]) - With no arguments, returns
the stop channel in MCS mode. Otherwise, sets the MCS stop
channel to
value.
mca_par("preset_passes" [,
value]) - With no arguments, returns
the current MCS-mode pass-preset value. Otherwise, sets the
number of passes.
mca_par("elapsed_passes") - Returns the
elapsed number of passes in MCS mode.
mca_par("current_chan") - In MCS mode,
returns the current channel of the current sweep.
mca_par("elapsed_dwell") - In MCS mode,
returns the elapsed dwell of the current channel of the current
sweep.
mca_par("mcs_xsync" [,
value]) - With no arguments, returns
0 if MCS-mode external sync is disabled or 1 if enabled.
Otherwise, sets the MCS external sync as specified by
value.
mca_par("pulse_source" [,
value]) - In MCS mode, with no
arguments, returns zero if the input pulses are taken from the
LLD/ULD filtered PHA input or nonzero if the input pulses are
taken as TTL pulses from the
XMCSPULSE input.
Otherwise, sets the pulse source according to
value.
mca_par("xmcsadvance_polarity" [,
"low"|"high"|0|1]) - With no arguments, returns 0 (for
active low) or 1 (for active high) to indicate the polarity of
the
XMCSADVANCE external channel advance input
signal. Otherwise, sets the polarity.
mca_par("xmcspulse_polarity" [,
"low"|"high"|0|1]) - With no arguments, returns 0 (for
active low) or 1 (for active high) to indicate the polarity of
the
XMCSPULSE MCS input signal. Otherwise, sets the
polarity.
mca_par("xmcsrestart_polarity" [,
"low"|"high"|0|1]) - With no arguments, returns 0 (for
active low) or 1 (for active high) to indicate the polarity of
the
XMCSRESTART MCS sweep advance input signal.
Otherwise, sets the polarity.